PROLINE Lyse2 5D

| Catalogue Number | Reagent Volume |
| 37001 | 1 x 500 mL |
Hematology analyzer reagent to destroy red blood cells (RBC) and then react with white blood cells (WBC) and hemoglobin (HGB).
Set the reagent temperature to room temperature, open the reagent package, insert the tube from the appropriate analyzer into the lyse bottle, and close the bottle cap. Replace the reagent according to the analyzer operator's manual, and then run the background to ensure that the measured value is within the background count range required in the analyzer operator's manual before testing the sample. See the analyzer operator's manual for more information.
Lyse is a solution that reacts with WBC which will separate itself into independent cells. WBC that reacts with lyse2 will maintain its cell morphology so that WBC can be classified into 5 types of cells, namely neutrophils, lymphocytes, monocytes, basophils and eosinophils.
- Stable & accurate linearity with a long shelf life
- Ready to use without analyzer adjustment
- Wide measurement range with high precision
- Equipped with a barcode for easy installation
- Cyanide-free reagent, safe & eco-friendly
| Blank value | WBC ≤0.2 × 10⁹/L |
| Reagent volume | 500 mL |
| Storage temperature | 2 – 30 °C |
| Operating temperature | 15 – 30 °C |
| Open vial stability | 60 days |
| Sample volume used (analyzer manual) | 20 µL |
| Expiration date | 24 months |
| Hematology 5 diff | PROLINE DF55 Auto Hematology Analyzer |
- Hematology Analyzer Reagent 5 diff
- Reagents bottle
- Brochure : INA
Contact our team to find out more product information and ordering
- Telp : +62 21 8984 2722
- WhatsApp : +62 815 1359 2626
- Email : marketing@proline.co.id
Contact our Technical support team for further assistance with product specifications, services and other technical documents.
- WA. +62 817 9324 884
- E. technical.support@proline.co.id