Urine Analyzer
A semi-quantitative instrument to analyze chemical content in urine using dry chemistry. The analysis process uses urine strips with options for 11 or 14 test parameters.
Innovatively designed, and portable. The PROLINE UA120 Urine Analyzer is connected to a barcode scanner, enabling quick and automated patient data entry with LIS, throughput of up to 240 tests/hour.
The device accurately measures up to 14 biochemical parameters, employing the colorimetric principle to detect biochemical constituents in urine. This is achieved through the color change that occurs when the reagent pad on the test strip reacts with the macroscopic biochemical components present in the urine sample.
Urobilinogen (URO) | Microalbumin (MA) |
Bilirubin (BIL) | Nitrite (NIT) |
Ketone (KET) | Leukocyte (LEU) |
Creatinine (Cr) | Glucose (GLU) |
Occult Blood (BLD) | pH |
Protein (PRO) | Vitamin C (VC) |
Specific Gravity (DG) | Urinary Calcium (Ca) |
- PROLINE URS14 Urine Strip 14 Parameters
- PROLINE URS11 Urine Strip 11 Parameters
Testing Procedure Using PROLINE UA120 Urine Analyzer
- Sample Preparation: Prepare the urine sample in a sample container.
- Dip the Strip: Submerge the PROLINE Urine Strip in the sample, ensuring all reagent pads are fully immersed. Remove the strip and shake off excess liquid inside the container.
- Clean the Strip: Wipe any excess sample with a dry cloth, avoiding contact with the reagent pads.
- Place the Strip: Position the PROLINE Urine Strip on the sample tray, ensuring the reagent pad end is at the top. Insert the urine strip; the analyzer will read it in 5 to 15 seconds and display the results on the monitor and thermal printer.
- Throughput of 60 tests/hour.
- Parameter options: 8, 9, 10, 11, 12, and 14.
- Stability CV < 1.0%.
- Operation with buttons, results displayed on a 240 x 148-pixel screen.
- Built-in thermal printer.
- Storage capacity of 10,000 records, LIS compatible.
- Reading time <10 seconds.
Feature | Description |
---|---|
Mode | Manual |
Method | Reflection photometry by color ratio analysis |
Detector | Photodiode |
Throughput | Normal:120 samples/hour Slow mode: 60 samples/hour |
Sample test | Single and continuous test |
Sample processing | routine, STAT and QC |
Time reading | 5 - 15 second |
Parameter | 8 - 14 parameters |
Capacity | 10.000 data |
Wavelength | 470 nm, 525 nm, 625 nm |
Light source | High-lumination LED |
Operating | Operated with a touchpad, the analysis results are displayed on a 240 x 148 pixel LCD |
Connectivity | 2 USB ports, 1 RS-232C Port (interface with PC external and Bi-directional LIS/HL7) |
Calibration Mode | Automatic calibration with calibration strips |
External Casing | Hard Plastic |
Printer | Internal thermal printer, paper size 57 mm |
Repeatability | CV≤1,0 % |
Stability | CV≤1,0 % |
Operating Requirement | Temperature: 20 – 25 °C Humidity: <80%, Atmosphere pressure: 76kPa - 106kPa |
Fuse | ф5mm×20mm, T2AL250V |
Power Supply | 100 – 230 VAC, 50 – 60 Hz, ≤45VA |
Dimension | 26,4 x 23,3 x 14,3 cm |
Weight | 2kg |
*HL7 connection is a global standard for transferring clinical and administrative data between applications.
Reach out to our team for more product and orders information.
- WA. +62 815 1359 2626
- E. marketing@proline.co.id
Contact our Technical Assistance team for further assistance with product specifications, services and other technical documents...
- WA. +62 817 9324 884
- E. technical.support@prodis.co.id