DYMIND DH33 Auto Hematology Analyzer

Reliable Features. Maximum Capacity.

Easy to operate with a reliable result.

DYMIND DH33 Auto Hematology Analyzer is manufactured locally by PROLINE (part of the Prodia Group which is engaged in the IVD medical device industry). These hematology instruments are manufactured under strict quality standards under license from Shenzhen Dymind Biotechnology Co., Ltd. and meet the requirements of GMPMD (Good Manufacturing Practices for Medical Devices). DYMIND DH33 Auto Hematology Analyzer provides 3 types of white blood cell reading technology features: granulocytes, monocytes, and lymphocytes as well as red blood cell and platelet count which provide convenient inspection for users. This relatively small-size hematology instrument is suitable for small to medium-scale laboratories.

26.4 cm TFT Touch Screen

Internal Cabinets for Lysis Solutions

Thermal Printer System

Distribution Permit No. KEMENKES RI AKD 20205320301
Model DYMIND DH33 Auto Hematology Analyzer
Throughput 60 samples/hour
Detection channel Dual-channel for counting
Leading technology Pulse baseline tracking technology, waveform discrimination technology, and
intelligent floating threshold technology
Detection principle Impedance method (WBC/RBC/PLT) and cyanide-free colorimetric method (HGB)
Detection systems The detection items include 21 parameters: report parameters (WBC, Lym#, Mid#, Gran#, Lym%, Mid%,
Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, and P-LCC)
, 3 histograms, alarm for abnormal erythrocytes, leukocyte, and platelet
Sample volume Minimum volume 9 µL
Counting modes Venous whole blood, capillary whole blood, and pre-diluted.
User interface Graphics are displayed on a 26.4 cm TFT touch screen, equipped with a thermal
printer with an additional printer connection and printout templates
Storage capacity 50,000 sample results, including parameter, histogram and patient information
Input/Output 4 USB ports and Bi-directional LIS connection
Operating environment Temperature 10℃ to 30℃; relative humidity ≤ 85%.
Power requirement AC 100 –240 V, 60/50 Hz.
Dimension 36,4 cm (P) x 41,7 cm (L) x 47,7 cm (T)
Weight ≤ 25 kg

Carryover Performance

Parameter Carryover
WBC ≤0,5 %
RBC ≤0,5 %
HGB ≤0,5 %
PLT ≤1 %
HCT ≤0,5 %

Display Range & Normal Background Performance

Parameter Linearity Range Display Range Normal Background
WBC (0 – 300)×10⁹ /L 0 – 999×10⁹ /L ≤0,2×10⁹ /L
RBC (0,00 – 8,50)×1012 /L 0 –18,00×10⁹ /L ≤0,02×1012 /L
HGB (0 – 250) g/L 0 – 300 g/L ≤1 g/L
PLT (0 – 3.000)×10⁹ /L 0 – 5.000×10⁹ /L ≤5×10⁹ /L
HCT (0 – 67) % 0 – 300 % ≤0,5 %

Linearity Range Performance

Parameter Linearity Range Deviation Range
WBC (0,00 – 100,00)×10⁹ /L ±0,30×10⁹ /L or ±5 %
(100,01 – 300,00)×10⁹ /L ±10 %
RBC (0,00 – 8,50)×1012 /L ±0,05×1012 /L or ±5 %
HGB (0 – 250) g/L ±2 g/L or ±2 %
PLT (0 – 1.000)×10⁹ /L ** ±10×10⁹ /L or ±8 %
(1.001– 3.000)×10⁹ /L ** ±12 %
HCT (0 – 67) % ±2 % (HCT value) or ±3 % (deviation percentage)

*Whole blood mode ** RBC≤7,0

Repeatability Performance

Parameter Condition Repeatability (CV%/absolute deviation d * )
WBC 7,0 – 15,0)×10⁹ /L ≤2,0 %
(4,0 – 6,9)×10⁹ /L ≤3,5 %
RBC (3,5 – 6,5)×1012 /L ≤1,5 %
HGB (100 - 180) g/L ≤1,5 %
MCV (70 – 110) fL ≤0,5 %
PLT (100 – 149)×10⁹ /L ≤5,0 %
(150– 500)×10⁹ /L ≤4,0 %

These repeatability requirements apply only to the situation in which a qualified sample has been run 11 times and the results of the 2nd to 11th runs are used to calculate the repeatabilities.

* Absolute deviation d = analysis result – an average of analysis results

Order Information