PROLINE HDL Precipitant
A diagnostic reagent for the quantitative examination of high-density lipoprotein cholesterol (HDL-C) in human serum, heparin plasma, or EDTA plasma in vitro using a photometric system.
HDL precipitant is an auxiliary reagent used with a cholesterol reagent to measure HDL levels in a sample. Chylomicrons, VLDL, and LDL are precipitated using phosphotungstic acid and magnesium ions. After centrifugation, the remaining supernatant contains HDL, which is then enzymatically analyzed.
Mix 200 µL of the sample with 500 µL of HDL precipitant reagent and incubate for 15 minutes at room temperature. After incubation, centrifuge the mixture for 20 minutes at 2500 g. Once centrifugation is complete, take 100 µL of the supernatant and mix it with 1000 µL of the cholesterol reagent. Incubate for another 10 minutes, then measure the result using an analyzer.
The precipitation of chylomicrons (VLDL and LDL) is achieved by adding phosphotungstic acid and magnesium ions to the sample. HDL Precipitant reagent is used in conjunction with the Cholesterol reagent.
- Ready-to-use liquid reagent in an open-system format, requiring no reconstitution.
- Guaranteed functional testing and reagent installation.
- Excellent linearity and stability performance.
- Available in MPK (Multi-Purpose Kit) bottle type.
- Compatible with over 65 brands of manual clinical chemistry analyzers.
Sample type | Human serum, heparin plasma or EDTA plasma. |
Measurement range | 2 mg/dL - 400 mg/dL |
Analysis wavelength | 546 nm |
Analysis mode | End-Point |
Reagent volume used (analyzer manual) | 500 µL |
Sample volume used (analyzer manual) | 200 µL |
Storage temperature | 15 - 25 °C |
Open vial stability | 22 months |
Expiration date | 22 months |
Catalogue Number | R/Mono Reagent Volume |
1 3540 99 90 019 | 2 x 50 mL |
1 3540 99 90 020 | 4 x 25 mL |
1 3540 99 90 022 | 6 x 20 mL |
1 3540 99 90 024 | 5 X 50 mL |
Reference Range: | |
According to the National Cholesterol Education Program (NCEP) guidelines | |
Low HDL cholesterol (major risk factor for coronary heart disease) | <40 mg/dL (<1.04 mmol/L) |
High HDL cholesterol (negative risk factor for coronary heart disease) | ≥60 mg/dL (≥1.55 mmol/L) |
- High-Density Lipoprotein Reagent
- Doos
- Kit insert
- Reagents bottle
- PROLINE. PROLINE HDL Precipitant Kit Insert [Update: 13540 01–Feb2025/02]
- Aterosklerosis
- Metabolisme Lipid
Contact our team to find out more about product information and ordering.
- Telp. +62 21 8984 2722
- WhatsApp. +62 815 1359 2626
- Surel. marketing@proline.co.id
Contact our Technical support team for further assistance with product specifications, services, and other technical documents.
- Telp. +62 21 8984 2722
- WhatsApp. +62 817 9324 884
- Surel. technical.support@prodis.co.id